An isoperibol drop calorimeter

Autor(en): Blachnik, R
Besser, J
Wallbrecht, P
Dreyer, K
Stichwörter: calorimetry; Chemistry; Chemistry, Analytical; Chemistry, Physical; drop calorimeter; heat content; Thermodynamics
Erscheinungsdatum: 1996
Herausgeber: ELSEVIER SCIENCE BV
Journal: THERMOCHIMICA ACTA
Volumen: 271
Startseite: 85
Seitenende: 92
Zusammenfassung: 
An isoperibol calorimeter is described. Its operation involves the drop of a sample from temperatures up to 1373 K into a metal block at 298.15 K. By this method the enthalpy contents of the samples can be measured. The calibration factor, obtained by electrical and drop calibrations, had a reproducibility of /-0.11%. The calorimeter was tested by determining the enthalpy of silver; agreement with the data of Barin was within a range of 1%.
Beschreibung: 
11th Ulm Conference on Developments in Calorimetry 1995, TECH UNIV BERGAKADEMIE BREIBERG, FREIBERG, GERMANY, MAR 29-31, 1995
ISSN: 00406031
DOI: 10.1016/0040-6031(95)02587-1