An isoperibol drop calorimeter

DC FieldValueLanguage
dc.contributor.authorBlachnik, R
dc.contributor.authorBesser, J
dc.contributor.authorWallbrecht, P
dc.contributor.authorDreyer, K
dc.date.accessioned2021-12-23T16:06:53Z-
dc.date.available2021-12-23T16:06:53Z-
dc.date.issued1996
dc.identifier.issn00406031
dc.identifier.urihttps://osnascholar.ub.uni-osnabrueck.de/handle/unios/7597-
dc.description11th Ulm Conference on Developments in Calorimetry 1995, TECH UNIV BERGAKADEMIE BREIBERG, FREIBERG, GERMANY, MAR 29-31, 1995
dc.description.abstractAn isoperibol calorimeter is described. Its operation involves the drop of a sample from temperatures up to 1373 K into a metal block at 298.15 K. By this method the enthalpy contents of the samples can be measured. The calibration factor, obtained by electrical and drop calibrations, had a reproducibility of /-0.11%. The calorimeter was tested by determining the enthalpy of silver; agreement with the data of Barin was within a range of 1%.
dc.language.isoen
dc.publisherELSEVIER SCIENCE BV
dc.relation.ispartofTHERMOCHIMICA ACTA
dc.subjectcalorimetry
dc.subjectChemistry
dc.subjectChemistry, Analytical
dc.subjectChemistry, Physical
dc.subjectdrop calorimeter
dc.subjectheat content
dc.subjectThermodynamics
dc.titleAn isoperibol drop calorimeter
dc.typeconference paper
dc.identifier.doi10.1016/0040-6031(95)02587-1
dc.identifier.isiISI:A1996TV57400011
dc.description.volume271
dc.description.startpage85
dc.description.endpage92